Exercise: Measurements And Instrumentation

Questions for: Measurements And Instrumentation

Assertion (A): Thermocouple instruments can be used for ac only.

Reason (R): Thermocouple instrument has a non linear scale.

A:
Both A and R are true and R is correct explanation of A
B:
Both A and R are true but R is not correct explanation of A
C:
A is true R is false
D:
A is false R is true
Answer: D

Thermocouple Instrument can be used for both ac and dc.

The percentage limiting error in an instrument reading 8.3 V and having guaranteed accuracy of 1% of the full scale reading and range of 0-150 V is
A:
1.81%
B:
0.181%
C:
18.1%
D:
0.0018%
Answer: C

Error = 1% of 150 = 1.5 V,

.

If the secondary winding of CT is opened when primary coil is carrying current
A:
transformer will burn immediately
B:
there will be weak flux density in core
C:
there will be very high induced voltage in secondary
D:
there will be high current in secondary
Answer: C

Since the primary of CT always remains excited, a high voltage is induced in secondary if the secondary is opened.

Assertion (A): In loss of charge method, it is preferable to use an electronic voltmeter.

Reason (R): The resistance of electronic voltmeter is very high.

A:
Both A and R are true and R is correct explanation of A
B:
Both A and R are true but R is not correct explanation of A
C:
A is true R is false
D:
A is false R is true
Answer: A

If resistance of voltmeter is not high, accuracy is affected.

It is required to study transients during switching process. Which CRO should be preferred?
A:
Ordinary CRO
B:
Dual beam CRO
C:
Dual trace CRO
D:
Storage oscilloscope
Answer: D

We can store the trace.